GC Analysis of Repeat Injections of a QC Test Mix Containing Water on Watercol™ 1910, 96 °C

column Watercol 1910, 30 m x 0.25 mm I.D., 0.20 µm (29711-U) oven 96 C carrier gas helium, 26 cm/sec sample 6 analytes in acetone, each at 1 mg/mL except water (at 0.5 mg/mL) injection 1 µL, 100:1 split liner 4 mm I.D., split type, cup design inj. temp. 250 °C detector TCD, 200 °C Analysis Note The ...

GC Analysis of Water in Ethanol (0.25%Standard) on Watercol™ 1460

column Watercol 1460, 30 m x 0.25 mm I.D., 0.20 µm; two identical columns installed in the same inlet, each going to a separate detector (29695-U) oven 45 °C (4 min), 5 °C/min to 75 °C, 10 °C/min to 220 °C (10 min) carrier gas helium, 20 cm/sec (measured in each column at an oven temperature of 135 ...

GC Analysis of a QC Test Mix Before/After Repeat 60%Water Injections on Watercol™ 1910

column Watercol 1910, 30 m x 0.25 mm I.D., 0.20 µm (29711-U) oven 163 °C carrier gas helium, 26 cm/sec sample 6 analytes in acetone, each at 1 mg/mL except water (at 0.5 mg/mL) injection 0.5 µL, 100:1 split liner 4 mm I.D., split type, cup design inj. temp. 250 °C detector TCD, 200 °C Analysis Note ...

GC Analysis of a QC Test Mix Containing Water on Watercol 1910, 96 °C

column Watercol 1910, 30 m x 0.25 mm I.D., 0.20 µm (29711-U) oven 96 C carrier gas helium, 26 cm/sec sample 6 analytes in acetone, each at 1 mg/mL except water (at 0.5 mg/mL) injection 1 µL, 100:1 split liner 4 mm I.D., split type, cup design inj. temp. 250 °C detector TCD, 200 °C Analysis Note The ...

GC Analysis of Water in Ethanol (0.25%Standard) on Watercol™ 1910

column Watercol 1910, 30 m x 0.25 mm I.D., 0.20 µm (29711-U) oven 45 °C (4 min), 5 °C/min to 75 °C, 10 °C/min to 190 °C carrier gas helium, 20 cm/sec (measured in each column at an oven temperature of 135 °C) sample water at 0.25% (v/v) in ethanol injection 1 µL, 50:1 split liner 4 mm I.D., split/sp...

GC Analysis of a QC Test Mix Containing Water on Watercol™ 1460, 110 °C

column Watercol 1460, 30 m x 0.25 mm I.D., 0.20 µm (29695-U) oven 110 °C carrier gas helium, 26 cm/sec sample 6 analytes in acetone, each at 1 mg/mL except water (at 0.5 mg/mL) injection 1 µL, 100:1 split liner 4 mm I.D., split type, cup design inj. temp. 250 °C detector TCD, 200 °C Analysis Note Th...

GC Analysis of Water in Ethanol (0.5%Standard) on Watercol™ 1460

column Watercol 1460, 30 m x 0.25 mm I.D., 0.20 µm (29695-U) oven 60 °C (12 min) carrier gas helium, 26 cm/sec sample water at 0.05% (v/v) in ethanol injection 0.5 µL, 100:1 split liner 4 mm I.D., split type, cup design inj. temp. 250 °C detector TCD, 200 °C Analysis Note The "Watercol" series of io...

GC Analysis of a QC Test Mix Containing Water on Watercol 1460, 96 °C

column Watercol 1460, 30 m x 0.25 mm I.D., 0.20 µm (29695-U) oven 96 C carrier gas helium, 26 cm/sec sample 6 analytes in acetone, each at 1 mg/mL except water (at 0.5 mg/mL) injection 1 µL, 100:1 split liner 4 mm I.D., split type, cup design inj. temp. 250 °C detector TCD, 200 °C Analysis Note The ...

GC Analysis of Aroma Fusel Alcohols in Tennessee Whiskey on Watercol™ 1910

column Watercol 1910, 30 m x 0.25 mm I.D., 0.20 µm; two identical columns installed in the same inlet, each going to a separate detector (29711-U) oven 50 °C (1 min), 5 °C/min to 180 °C (3 min) carrier gas helium, 25 cm/sec (measured in each column at an oven temperature of 125 °C) sample neat Tenne...

GC Analysis of Water (0.05-1%Calibration Curve) on Watercol™ 1460

column Watercol 1460, 30 m x 0.25 mm I.D., 0.20 µm (29695-U, two identical columns installed in the same inlet, each going to a separate detector) oven 45 °C (4 min), 5 °C/min to 75 °C, 10 °C/min to 220 °C (10 min) carrier gas helium, 20 cm/sec (measured in each column at an oven temperature of 135 ...

GC Analysis of Water Impurity in C1-C6 Alcohols on Watercol™ 1910

column Watercol 1910, 30 m x 0.25 mm I.D., 0.20 µm (29711-U) oven 40 °C (2 min), 4 °C/min to 125 °C (1 min) carrier gas helium, 1.1 mL/min sample 10-component alcohol mix in methylene chloride, 500 µg/mL total concentration (mixture adsorbed some water during storage) injection 1 µL, 100:1 split lin...

GC Analysis for Aroma Fusel Alcohols in Vodka on Watercol™ 1910

column Watercol 1910, 30 m x 0.25 mm I.D., 0.20 µm; two identical columns installed in the same inlet, each going to a separate detector (29711-U) oven 50 °C (1 min), 5 °C/min to 180 °C (3 min) carrier gas helium, 25 cm/sec (measured in each column at an oven temperature of 125 °C) sample neat vodka...

GC Analysis of Water in Ethanol (0.5%Standard) on Watercol™ 1900

column Watercol 1900, 30 m x 0.25 mm I.D., 0.20 µm (29761-U) oven 80 °C (10 min) carrier gas helium, 26 cm/sec sample water at 0.05% (v/v) in ethanol injection 0.5 µL, 100:1 split liner 4 mm I.D., split type, cup design inj. temp. 250 °C detector TCD, 200 °C Analysis Note The "Watercol" series of io...

GC Analysis of Water in Ethanol (0.05%Standard) on Watercol™ 1900

column Watercol 1900, 30 m x 0.25 mm I.D., 0.20 µm (29761-U) oven 80 °C (10 min) carrier gas helium, 26 cm/sec sample water at 0.05% (v/v) in ethanol injection 0.5 µL, 100:1 split liner 4 mm I.D., split type, cup design inj. temp. 250 °C detector TCD, 200 °C Analysis Note The "Watercol" series of io...

GC Analysis of a QC Test Mix Before/After Repeat 100%Water Injections on Watercol™ 1460

column Watercol 1460, 30 m x 0.25 mm I.D., 0.20 µm (29695-U) oven 110 °C carrier gas helium, 26 cm/sec sample 6 analytes in acetone, each at 1 mg/mL except water (at 0.5 mg/mL) injection 1 µL, 100:1 split liner 4 mm I.D., split type, cup design inj. temp. 250 °C detector TCD, 200 °C Analysis Note Th...

GC Analysis of a QC Test Mix Containing Water on Watercol 1900, 96 °C

column Watercol 1900, 30 m x 0.25 mm I.D., 0.20 µm (29761-U) oven 96 C carrier gas helium, 26 cm/sec sample 6 analytes in acetone, each at 1 mg/mL except water (at 0.5 injection 1 µL, 100:1 split liner 4 mm I.D., split type, cup design inj. temp. 250 °C detector TCD, 200 °C Analysis Note The "Waterc...

GC Analysis of Water in Ibuprofen (Spiked) on Watercol™ 1910

column Watercol 1910, 30 m x 0.25 mm I.D., 0.20 µm (29711-U) oven 35 °C (4 min), 8 °C/min to 130 °C (1 min) carrier gas helium, 11.40 psi constant pressure (equal to 25 cm/sec), measured with oven at 35 °C sample 0.198 g ibuprofen weighed into a 1 mL volumetric flask, 0.5 mL acetone added, vortexed ...

GC Analysis of Aroma Fusel Alcohols in Rum on Watercol™ 1910

column Watercol 1910, 30 m x 0.25 mm I.D., 0.20 µm; two identical columns installed in the same inlet, each going to a separate detector (29711-U) oven 50 °C (1 min), 5 °C/min to 180 °C (3 min) carrier gas helium, 25 cm/sec (measured in each column at an oven temperature of 125 °C) sample neat rum i...

GC Analysis of Water in Gasoline (Unspiked) on Watercol™ 1460

column Watercol 1460, 30 m x 0.25 mm I.D., 0.20 µm; two identical columns installed in the same inlet, each going to a separate detector (29695-U) oven 45 °C (4 min), 5 °C/min to 75 °C, 10 °C/min to 220 °C (10 min) carrier gas helium, 20 cm/sec (measured in each column at an oven temperature of 135 ...

GC Analysis of Water in Gasoline (Spiked) on Watercol™ 1460

column Watercol 1460, 30 m x 0.25 mm I.D., 0.20 µm; two identical columns installed in the same inlet, each going to a separate detector (29695-U) oven 45 °C (4 min), 5 °C/min to 75 °C, 10 °C/min to 220 °C (10 min) carrier gas helium, 20 cm/sec (measured in each column at an oven temperature of 135 ...

GC/BID Analysis of Aromatics and Water in Solvents on Watercol™ 1910

column Watercol™ 1910 30m x 0.25mm x 0.20 µm (29711-U) oven 40°C (10 minutes) 8°C/minute to 170°C (5 minutes) carrier gas helium, 25 cm/sec sample BTEX spiked with 20ul of di-water in Hexane injection 1 µL, 100:1 split liner Split with wool inj. temp. 250 °C detector BID, 200 °C Analysis Note Easily...

GC Analysis of an 8-Component Solvent Mix in Water on Watercol™ 1910

column Watercol 1910, 30 m x 0.25 mm I.D., 0.20 µm (29711-U) oven 35 °C, 4 °C/min to 125 °C (2 min) carrier gas helium, 25 cm/sec sample 8-component solvent mix in water injection 1 µL, 100:1 split liner 4 mm I.D., split type, cup design inj. temp. 250 °C detector TCD, 300 °C Analysis Note The "Wate...

GC Analysis of Water (0.1-100%Calibration Curve) on Watercol™ 1460

column Watercol 1460, 30 m x 0.25 mm I.D., 0.20 µm (29695-U) oven 60 °C (12 min) carrier gas helium, 26 cm/sec sample 7 standards, water at 0.1, 0.5, 20, 40, 60, and 80% (v/v), each in ethanol, and water at 100% injection 0.5 µL for 0.1-60% water, 0.1 µL for 80-100% water, 100:1 split liner 4 mm I.D...

GC Analysis of Water in a 60-Component Solvent Mix on Watercol™ 1910

column Watercol 1910, 30 m x 0.25 mm I.D., 0.20 µm; two identical columns installed in the same inlet, each going to a separate detector (29711-U) oven 50 °C (1 min), 5 °C/min to 180 °C (5 min) carrier gas helium, 25 cm/sec (measured in each column at an oven temperature of 125 °C) sample mixture of...

GC Analysis of Aroma Fusel Alcohols in Brandy on Watercol™ 1910

column Watercol 1910, 30 m x 0.25 mm I.D., 0.20 µm ; two identical columns installed in the same inlet, each going to a separate detector (29711-U) oven 50 °C (1 min), 5 °C/min to 180 °C (3 min) carrier gas helium, 25 cm/sec (measured in each column at an oven temperature of 125 °C) sample neat bran...

GC Analysis of Water in Ethanol (0.5%Standard) on Watercol™ 1910

column Watercol 1910, 30 m x 0.25 mm I.D., 0.20 µm (29711-U) oven 80 °C (10 min) carrier gas helium, 26 cm/sec sample water at 0.05% (v/v) in ethanol injection 0.5 µL, 100:1 split liner 4 mm I.D., split type, cup design inj. temp. 250 °C detector TCD, 200 °C Analysis Note The "Watercol" series of io...

GC Analysis of Repeat Injections of a QC Test Mix Containing Water on Watercol™ 1900, 96 °C

column Watercol 1900, 30 m x 0.25 mm I.D., 0.20 µm (29761-U) oven 96 C carrier gas helium, 26 cm/sec sample 6 analytes in acetone, each at 1 mg/mL except water (at 0.5 mg/mL) injection 1 µL, 100:1 split liner 4 mm I.D., split type, cup design inj. temp. 250 °C detector TCD, 200 °C Analysis Note The ...

GC Analysis of Water in Ethanol (0.05%Standard) on Watercol™ 1910

column Watercol 1910, 30 m x 0.25 mm I.D., 0.20 µm (29711-U) oven 80 °C (10 min) carrier gas helium, 26 cm/sec sample water at 0.05% (v/v) in ethanol injection 0.5 µL, 100:1 split liner 4 mm I.D., split type, cup design inj. temp. 250 °C detector TCD, 200 °C Analysis Note The "Watercol" series of io...

GC Analysis of a QC Test Mix Before/After Repeat 60%Water Injections on Watercol™ 1460

column Watercol 1460, 30 m x 0.25 mm I.D., 0.20 µm (29695-U) oven 110 °C carrier gas helium, 26 cm/sec sample 6 analytes in acetone, each at 1 mg/mL except water (at 0.5 mg/mL) injection 1 µL, 100:1 split liner 4 mm I.D., split type, cup design inj. temp. 250 °C detector TCD, 200 °C Analysis Note Th...

GC Analysis of a QC Test Mix Containing Water on Watercol 1910, 163 °C

column Watercol 1910, 30 m x 0.25 mm I.D., 0.20 µm (29711-U) oven 163 °C carrier gas helium, 26 cm/sec sample 6 analytes in acetone, each at 1 mg/mL except water (at 0.5 mg/mL) injection 0.5 µL, 100:1 split liner 4 mm I.D., split type, cup design inj. temp. 250 °C detector TCD, 200 °C Analysis Note ...